7.4 Calibration Settings
® Commissioning (Start-up)
® Calibration
Function Selection Factory setting
Hold during calibration Yes, No Yes
Entry of temperature coefficient Tc of 0.0 to 10.0 %/K 2.1 %/K
calibration solution
Entry of temperature of calibration 31 to +482°F (35 to +250°C) 77°F (25°C)
solution (MTC only)
7.5 Cleaning Function
® Commissioning (Start-up)
® Chemoclean
An instrument with a total of five output contacts is needed for the Chemoclean function
with a contacting measuring cell.
The holder CLA 611 and spray head CLR 30 are required to automate the cleaning
function.
The Chemoclean cleaning function is to be adjusted in the System set-up / Output
relays (see chapter 7.1.4).
Chemoclean is started from the Short-cut to relays menu (see chapter 6.7).
Note!
The options Chemoclean off and Setting cannot be accessed during an
ongoing cleaning cycle.
The cleaning cycle starts again from the beginning when changes have been
made in the Settings menu.