7.4 Calibration Settings
® Commissioning (Start-up)
® Calibration
Function Selection Factory setting
Hold during calibration Yes, No Yes
Entry of temperature coefficient Tc of 0.0 to 10.0 %/K 2.1 %/K
calibration solution
Entry of temperature of calibration 31 to +482°F (35 to +250°C) 77°F (25°C)
solution (MTC only)
7.5 Cleaning Function
® Commissioning (Start-up)
® Chemoclean
An instrument with a total of five output contacts is needed for the Chemoclean function
with a contacting measuring cell.
The holder CLA 611 and spray head CLR 30 are required to automate the cleaning
function.
The Chemoclean cleaning function is to be adjusted in the System set-up / Output
relays (see chapter 7.1.4).
Chemoclean is started from the Short-cut to relays menu (see chapter 6.7).
Note!
The options Chemoclean off and Setting cannot be accessed during an
ongoing cleaning cycle.
The cleaning cycle starts again from the beginning when changes have been
made in the Settings menu.

Table of Contents

Other manuals for Endress+Hauser Mycom CLM 152

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Endress+Hauser Mycom CLM 152 and is the answer not in the manual?

Endress+Hauser Mycom CLM 152 Specifications

General IconGeneral
BrandEndress+Hauser
ModelMycom CLM 152
CategoryTransmitter
LanguageEnglish

Related product manuals