PROline Promag 50 10 Technical data
Endress+Hauser 107
10 Technical data
10.1 Technical data at a glance
10.1.1 Application
Application Measuring the flow rate of fluids in closed piping systems.
A minimum conductivity of 5 µS/cm is required for measuring; the minimum
conductivity required in the case of demineralised water is 20 µS/cm.
Applications in measuring, control and regulation technology
Liner specific applications:
Promag W (DN 25...2000):
Polyurethane lining for applications with cold water and for slightly abrasive fluids.
Hard rubber lining for all applications with water (especially for trinking water)
Promag P (DN 15...600):
PTFE lining for standard applications in chemical and process industries.
PFA lining for all applications in chemical and process industries; especially for high
process temperatures and applications with temperature shocks.
Promag H (DN 2...100):
PFA lining for all applications in chemical, process and food industries; especially for
high process temperatures, for applications with temperature shocks and for applica-
tions with CIP or SIP cleaning processes.
10.1.2 Function and system design
Measuring principle Electromagnetic flow measurement on the basis of Faraday’s Law.
Measuring system The measuring system consists of a transmitter and a sensor.
Two versions are available:
Compact version: transmitter and sensor form a single mechanical unit.
Remote version: transmitter and sensor are installed separately.
Transmitter :
•Promag 50
Sensor:
Promag W (DN 25...2000)
Promag P (DN 15...600)
Promag H (DN 2...100)
10.1.3 Input
Measured variable Flow rate (proportional to induced voltage)
Measuring range Typically v = 0.01…10 m/s with the specified measuring accuracy
Operable flow range Over 1000 : 1
Input signals Status input (auxiliary input):
U = 3…30 V DC, R
i
= 5 k, galvanically isolated
Configurable for: totalizer reset, positive zero return