6 Commissioning PROline Promag 50
82 Endress+Hauser
6.2.4 Current output: active/passive
The current output is configured as “active” or “passive” by means of various jumpers
on the I/O board.
#
Warning!
Risk of electric shock. Exposed components carry dangerous voltages. Make sure
that the power supply is switched off before you remove the cover of the electronics
compartment.
Wait at least 10 minutes for heat to dissipate before opening an Ex-rated device.
1. Switch off power suppy.
2. Remove the I/O board Page 97, 99
3. Position the jumper according to Fig. 46.
" Caution!
Risk of destroying the measuring device. Set the jumpers exactly as shown in
Fig. 46. Incorrectly set jumpers can cause overcurrents that would destroy either
the measuring device or external devices connected to it.
4. Installation of the I/O board is the reverse of the removal procedure.
Fig. 46: Configuring the current output (I/O board)
1 Active current output (Factory setting)
2 Passive current output
6.3 Data storage device (S-DAT)
The S-DAT is an exchangeable data storage device in which all sensor relevant param-
eters are stored, i.e., diameter, serial number, calibration factor, zero point.