PROline Promag 50 9Trouble-shooting
Endress+Hauser 87
9 Trouble-shooting
9.1 Trouble-shooting instructions
Always start trouble-shooting with the checklist below, if faults occur after start-up or
during operation. The routine takes you directly to the cause of the problem and the
appropriate remedial measures.
Check the display
No display visible and no
output signals present.
1. Check the power supply terminals 1, 2
2. Check the power line fuse
Page 101
85...260 V AC: 0.8 A slow-blow / 250 V
20...55 V AC and 16...62 V DC: 2 A slow-blow / 250 V
3. Measuring electronics defective
order spare parts Page 96
No display visible, but out-
put signals are present.
1. Check whether the ribbon-cable connector of the display module is
correctly plugged into the amplifier board
Page 98, 100
2. Display module defective
order spare parts Page 96
3. Measuring electronics defective
order spare parts Page 96
Display texts are in a foreign
language.
Switch off power supply. Press and hold down both the OS buttons and
switch on the measuring device. The display text will appear in English
(default) and is displayed at maximum contrast.
Measured value indicated,
but no signal at the current
or pulse output
Measuring electronics defective
order spare parts Page 96
Error messages on display
Errors which occur during commissioning or measuring operation are displayed immediately.
Error messages consist of a variety of icons: the meanings of these icons are as follows (example):
Error type: S = system error, P = process error
Error message type:
$ = fault message, ! = notice message
EMPTY PIPE = Type of error, e.g. measuring tube is only partly filled or completely empty
03:00:05 = duration of error occurrence (in hours, minutes and seconds)
# 401 = error number
" Caution!
See the information on Page 64!
The measuring system interprets simulations and positive zero return as system errors, but displays
them as notice message only.
Error number:
No. 001 – 399
No. 501 – 699
System error (device error) has occurred
Page 88
Error number:
No. 401
499
Process error (application error) has occurred
Page 92
Other error (without error message)
Some other error has
occurred.
Diagnosis and rectification Page 93