PROline Promag 50 10 Technical data
Endress+Hauser 111
Process
Medium temperature
range
The permissible fluid temperature depends on the lining of the measuring tube:
Promag W
0…+80 °C for hard rubber (DN 65…2000)
–20…+50 °C for polyurethane (DN 25…2000)
Promag P
–40…+130 °C for PTFE (DN 15…600), for restrictions → refer to diagrams
–20…+180 °C for PFA (DN 25…200), for restrictions → refer to diagrams
Fig. 53: Compact version Promag P (with PFA or PTFE lining)
T
A
= ambient temperature, T
F
= fluid temperature, HT = high-temperature version, with insulation
Fig. 54: Remote version Promag P (with PFA or PTFE lining)
T
A
= ambient temperature, T
F
= fluid temperature, HT = high-temperature version, with insulation
0
0
-20-40 20
20
40
60
T
A
[°C]
40 60 80
T
F
[°C]
100 120 140 160 180
HT
PTFE
PTFE
PFA
PFA
F06-5xPxxxxx-05-xx-xx-xx-000
0
0
-20-40 20
20
40
60
T
A
[°C]
40 60 80
T
F
[°C]
100 120 140 160 180
HT
PTFE
PTFE
PFA
PFA
F06-xxPxxxxx-05-xx-xx-xx-000