PROline Promag 50 10 Technical data
Endress+Hauser 121
Promag P
Nominal diameter
Pressure rating Inside diameter of
measuring tube
[mm] [inch]
EN (DIN)
[bar]
ANSI
[lbs]
JIS with PFA
[mm]
with PTFE
[mm]
15 1/2" PN 40 Cl 150 20K
−
15
25 1" PN 40 Cl 150 20K 23 26
32
−
PN 40
−
20K 32 35
40 1 1/2" PN 40 Cl 150 20K 36 41
50 2" PN 40 Cl 150 10K 48 52
65
−
PN 16
−
10K 63 67
80 3" PN 16 Cl 150 10K 75 80
100 4" PN 16 Cl 150 10K 101 104
125
−
PN 16
−
10K 126 129
150 6" PN 16 Cl 150 10K 154 156
200 8" PN 10 Cl 150 10K 201 202
250 10" PN 10 Cl 150 10K
−
256
300 12" PN 10 Cl 150 10K
−
306
350 14" PN 10 Cl 150
−−
337
400 16" PN 10 Cl 150
−−
387
450 18" PN 10 Cl 150
−−
432
500 20" PN 10 Cl 150
−−
487
600 24" PN 10 Cl 150
−−
593
Promag H
Nominal diameter
Pressure ratings * Inside diameter of
measuring tube **
[mm] [inch] [bar]
PFA
2 1/12" PN 16 / PN 40 2.25
4 5/32" PN 16 / PN 40 4.5
8 5/16" PN 16 / PN 40 9.0
15 1/2" PN 16 / PN 40 16.0
– 1" PN 16 / PN 40 22.6
25 – PN 16 / PN 40 26.0
40 1 1/2" PN 16 / PN 40 35.3
50 2" PN 16 / PN 40 48.1
65 2 1/2" PN 16 / PN 40 59.9
80 3" PN 16 / PN 40 72.6
100 4" PN 16 / PN 40 97.5
* The pressure rating depends on the process connections and the seals (see Page 112).
** Inside diameters of process connections
→ see Page 137 ff.