Device functions Proline Promag 55 FOUNDATION Fieldbus 3 Transducer Block
Endress+Hauser 89
Simulation -
Value Measurand
AUTO - OOS Use this parameter to specify a selectable value (e.g. 12 m³/s). This is used
to test the associated parameters in the device itself and downstream
signal loops.
User input:
5-digit floating-point number
Factory setting:
0 [unit]
! Note!
The unit is taken from the "System Unit - Volume Flow" or "System Unit -
Mass Flow" parameter ( → Page 79 ff.).
" Caution!
The setting is not saved if the power supply fails.
Simulation - Unit read only Use this parameter to display the current unit for the simulation value in
the "Simulation - Measurand Value" parameter.
! Note!
The unit can be selected in the "System Unit - Volume Flow" or "System
Unit - Mass Flow" parameter ( → Page 79).
Service/Analys. -
Measuring Period
read only This parameter is only used in the event of a service call.
Service/Analys. -
Risetime
read only This parameter is only used in the event of a service call.
Service/Analys. -
Electrode Pot. 1
read only This parameter is only used in the event of a service call.
Service/Analys. -
Electrode Pot. 2
read only This parameter is only used in the event of a service call.
Sensor - Type read only Use this function to view the sensor type.
Sensor -
SW Rev.No.S-DAT
read only Use this function to view the software revision number of the software
used to create the content of the S-DAT.
Sensor -
HW Rev.Number.
read only Use this parameter to view the hardware revision number of the sensor.
Sensor -
HW Identification
read only Use this parameter to view the hardware ID number of the sensor.
Sensor -
Prod.Number
read only Use this parameter to view the production number of the sensor.
Amp. Device Type read only This parameter is only used in the event of a service call.
"Flow" Transducer Block/base index 1400
Parameter
Write access with
operating mode
(MODE_BLK)
Description