Proline Promag W 500
Endress+Hauser 45
Performance characteristics
Reference operating
conditions
Error limits following DIN EN 29104, in future ISO 20456
Water, typically: +15 to +45 °C (+59 to +113 °F); 0.5 to 7 bar (73 to 101 psi)
Data as indicated in the calibration protocol
Accuracy based on accredited calibration rigs according to ISO 17025
Maximum measured error Error limits under reference operating conditions
o.r. = of reading
Volume flow
±0.5 % o.r. ± 1 mm/s (0.04 in/s)
Optional: ±0.2 % o.r. ± 2 mm/s (0.08 in/s)
Fluctuations in the supply voltage do not have any effect within the specified range.
2.5
[%]
2.0
1.5
1.0
0.5
0
0.2 %
0.5 %
0
1
2 4 6 8 10 [m/s]
v
5 10 15 20 25 30 32 [ft/s]0
A0028974
 23 Maximum measured error in % o.r.
2.5
[%]
2.0
1.5
1.0
0.5
0
0.2 %
0.5 %
0
[m/s]
v
[ft/s]
v
0.2
v
0.5
0
v
max
A0017051
 24 Flat Spec in % o.r.
Flat Spec flow values0.5 %
Nominal diameter v
0.5
[mm] [in] [m/s] [ft/s]
25 to 600 1 to 24 0.5 1.64

Table of Contents