Proline Promag W 500
Endress+Hauser 53
Storage temperature
The storage temperature corresponds to the operating temperature range of the transmitter and the
sensor →  52.
Protect the measuring device against direct sunlight during storage in order to avoid unacceptably
high surface temperatures.
Select a storage location where moisture cannot collect in the measuring device as fungus or
bacteria infestation can damage the liner.
If protection caps or protective covers are mounted these should never be removed before
installing the measuring device.
Degree of protection
Transmitter
As standard: IP66/67, type 4X enclosure
When housing is open: IP20, type 1 enclosure
Display module: IP20, type 1 enclosure
Sensor
As standard: IP66/67, type 4X enclosure
Optionally available for order:
IP66/67, type 4X enclosure; fully welded, with protective varnish EN ISO 12944 C5-M. Suitable
for use in corrosive atmospheres.
IP68, type 6P enclosure; fully welded, with protective varnish as per EN ISO 12944 C5-M.
Suitable for permanent immersion in water ≤ 3 m (10 ft) or up to 48 hours at depths ≤
10 m (30 ft).
IP68, type 6P enclosure; fully welded, with protective varnish as per EN ISO 12944 Im1/Im2/
Im3. Suitable for permanent immersion in saline water ≤ 3 m (10 ft) or up to 48 hours at depths
≤ 10 m (30 ft) or in buried applications.
External WLAN antenna
IP67
Vibration resistance
Vibration, sinusoidal according to IEC 60068-2-6
2 to 8.4 Hz, 7.5 mm peak
8.4 to 2 000 Hz, 2 g peak
Vibration broad-band random, according to IEC 60068-2-64
10 to 200 Hz, 0.01 g
2
/Hz
200 to 2 000 Hz, 0.003 g
2
/Hz
Total: 2.70 g rms
Shock resistance
Shock, half-sine according to IEC 60068-2-27
6 ms 50 g
Shock resistance
Shock due to rough handling following IEC 60068-2-31
Mechanical load
Protect the transmitter housing against mechanical effects, such as shock or impact; the use of the
remote version is sometimes preferable.
Never use the transmitter housing as a ladder or climbing aid.
Electromagnetic
compatibility (EMC)
As per IEC/EN 61326 and NAMUR Recommendation 21 (NE 21)
Device version with PROFIBUS DP: Complies with emission limits for industry as per EN 50170
Volume 2, IEC 61784
The following applies for PROFIBUS DP: If baud rates > 1.5 MBaud, an EMC cable entry must be
used and the cable shield must continue as far as the terminal wherever possible.
Details are provided in the Declaration of Conformity.
Process
Medium temperature range
0 to +80 °C (+32 to +176 °F) for hard rubber, DN 50 to 2000 (2 to 78")
–20 to +50 °C (–4 to +122 °F) for polyurethane, DN 25 to 1200 (1 to 48")