Proline Promag W 500
46 Endress+Hauser
Flat Spec flow values0.2 %
Nominal diameter v
0.2
[mm] [in] [m/s] [ft/s]
25 to 600 1 to 24 1.5 4.92
Electrical conductivity
Max. measured error not specified.
Accuracy of outputs
The outputs have the following base accuracy specifications.
Current output
Accuracy ±5 µA
Pulse/frequency output
o.r. = of reading
Accuracy Max. ±50 ppm o.r. (over the entire ambient temperature range)
Repeatability
o.r. = of reading
Volume flow
Max. ±0.1 % o.r. ± 0.5 mm/s (0.02 in/s)
Electrical conductivity
Max. ±5 % o.r.
Influence of ambient
temperature
Current output
Temperature coefficient Max. 1 μA/°C
Pulse/frequency output
Temperature coefficient No additional effect. Included in accuracy.
Installation
No special measures such as supports etc. are necessary. External forces are absorbed by the
construction of the device.
Mounting location
h
A0029343