EasyManua.ls Logo

Endress+Hauser Proline Promag W 500 - Page 47

Endress+Hauser Proline Promag W 500
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Proline Promag W 500
Endress+Hauser 47
Preferably install the sensor in an ascending pipe, and ensure a sufficient distance to the next pipe
elbow: h ≥ 2 × DN
Not necessary in the case of order code for "Design", option C "Insertion length short ISO/DVGW
to DN300, without inlet/outlet runs, constricted meas.tube"
1 1
A0033017
 25 Installation of the sensor after a control valve is not recommended
1 Control valve
Installation in down pipes
Install a siphon with a vent valve downstream of the sensor in down pipes whose length h ≥
5 m (16.4 ft). This precaution is to avoid low pressure and the consequent risk of damage to the
measuring tube. This measure also prevents the system losing prime.
h
2
1
A0028981
 26 Installation in a down pipe
1 Vent valve
2 Pipe siphon
h Length of down pipe
Installation in partially filled pipes
A partially filled pipe with a gradient necessitates a drain-type configuration.
2 x DN³
5 x DN³
A0029257
For heavy sensors DN ≥ 350 (14")
A0016276

Table of Contents

Related product manuals