Installation Proline Promag W 800
18 Endress+Hauser
Orientation
An optimum orientation position helps avoid gas and air accumulations and deposits in the
measuring tube.
Vertical orientation
Vertical orientation is optimal in the following scenarios:
For self-emptying piping systems.
For sludge containing sand or stones where the solids tend to settle at the bottom.
A0015591
Fig. 10: Vertical orientation
Horizontal orientation
The measuring electrode plane should be horizontal in the case of horizontal orientation. This
prevents brief insulation of the two measuring electrodes by entrained air bubbles.
A0016260
Fig. 11: Horizontal orientation
1 EPD electrode for the detection of empty pipes (not supported by the transmitter)
2 Measuring electrodes used for measuring signal pick up and empty pipe detection (EPD). An EPD alarm is triggered
if there is no fluid between electrodes.
3 Reference electrode for potential equalization