Connecting cable
In order to ensure measuring accuracy, comply with the following instructions when installing
a remote version:
• Fix the cable run or route it in an armored conduit. Cable movements can falsify the
measuring signal especially in the case of low fluid conductivities.
• Route the cable well clear of electrical machines and switching elements.
• Ensure potential equalization between sensor and transmitter, if necessary.
• The maximum connecting cable length is 20 m (35.6 ft).
GSM/GPRS antenna
Check the signal strength of the mobile communications network before mounting the
GSM/GPRS antenna.
Detailed information on checking the mobile communications network:
Operating Instructions for the device on the enclosed CD-ROM
6.2 Installing the measuring device
6.2.1 Installing the Promag W sensor
Required mounting tools
For flange and other process connections:
• Bolts, nuts, seals, etc. These items are not included in the scope of supply and must be supplied
by the customer.
• Appropriate mounting tools
Installing the sensor
Mount the sensor between the pipe flanges.
In doing so, please note the following:
• Required screw tightening torques → 23.
• If using ground disks:
Comply with the Installation Instructions provided with the ground disks.
Mounting the seals
"
Caution!
Risk of short circuit!
Do not use electrically conductive sealing compounds such as graphite! An electrically
conductive layer could form on the inside of the measuring tube and short-circuit the measuring
signal.
Comply with the following instructions when installing seals:
• For hard rubber lining: additional seals are always required.
• For polyurethane lining: generally additional seals are not required.
• For DIN flanges: only use seals according to EN 1514-1.