Diagnostics and troubleshooting Proline Promass I 300 PROFINET
202 Endress+Hauser
Diagnostic information Remedy instructions
No. Short text
537 Configuration 1. Check IP addresses in network
2. Change IP address
Measured variable status
Quality Good
Quality substatus Ok
Coding (hex) 0x80 to 0x83
Status signal F
Diagnostic behavior Warning
Influenced measured variables
–
Diagnostic information Remedy instructions
No. Short text
594 Relay output simulation Deactivate simulation switch output
Measured variable status
Quality Good
Quality substatus Ok
Coding (hex) 0x80 to 0x83
Status signal C
Diagnostic behavior Warning
Influenced measured variables
–
12.7.4 Diagnostic of process
Diagnostic information Remedy instructions
No. Short text
803 Current loop 1. Check wiring
2. Change I/O module
Measured variable status
Quality Good
Quality substatus Ok
Coding (hex) 0x80 to 0x83
Status signal F
Diagnostic behavior Alarm
Influenced measured variables
–