Proline Promass I 300 PROFINET Technical data
Endress+Hauser 247
The minimum recommended full scale value is approx. 1/20 of the maximum full scale
value
In most applications, 20 to 50 % of the maximum full scale value can be considered ideal
A low full scale value must be selected for abrasive media (such as liquids with entrained
solids): flow velocity < 1 m/s (< 3 ft/s).
For gas measurement the following rules apply:
The flow velocity in the measuring tubes should not exceed half the sound velocity
(0.5 Mach).
The maximum mass flow depends on the density of the gas: formula →  230
To calculate the flow limit, use the Applicator sizing tool →  228
Pressure loss To calculate the pressure loss, use the Applicator sizing tool →  228
System pressure →  24
16.10 Mechanical construction
Design, dimensions
For the dimensions and installation lengths of the device, see the "Technical
Information" document, "Mechanical construction" section.
Weight All values (weight exclusive of packaging material) refer to devices with EN/DIN
PN 40 flanges. Weight specifications including transmitter as per order code for "Housing",
option A "Aluminum, coated".
Different values due to different transmitter versions:
Transmitter version for the hazardous area
(Order code for "Housing", option A "Aluminum, coated"; Ex d): +2 kg (+4.4 lbs)
Cast transmitter version, stainless
(Order code for "Housing", option L "Cast, stainless"): +6 kg (+13 lbs)
Transmitter version for hygienic area
(Order code for "Housing", option B "Stainless, hygienic"): +0.2 kg (+0.44 lbs)
Weight in SI units
DN
[mm]
Weight [kg]
8 11
15 13
15 FB 19
25 20
25 FB 39
40 40
40 FB 65
50 67
50 FB 118
80 122
FB = Full bore