System integration Proline Promass I 300 PROFINET
86 Endress+Hauser
Liquid type Enum code
Ammonium nitrate in water 13
Iron(III) chloride in water 14
% mass / % volume 19
User Profile Coef Set No. 1 21
User Profile Coef Set No. 2 22
User Profile Coef Set No. 3 23
9.3.3 Status coding
Status Coding (hex) Meaning
BAD - Maintenance alarm 0x24 A measured value is not available because a device
error has occurred.
BAD - Process related 0x28 A measured value is not available because the process
conditions are not within the device's technical
specification limits.
BAD - Function check 0x3C A function check is active (e.g. cleaning or calibration)
UNCERTAIN - Initial value 0x4F A pre-defined value is output until a correct measured
value is available again or until remedial measures
have been carried out that change this status.
UNCERTAIN -
Maintenance demanded
0x68 Signs of wear and tear have been detected on the
measuring device. Short-term maintenance is needed
to ensure that the measuring device remains
operational.
The measured value might be invalid. The use of the
measured value depends on the application.
UNCERTAIN - Process
related
0x78 The process conditions are not within the device's
technical specification limits. This could have a
negative impact on the quality and accuracy of the
measured value.
The use of the measured value depends on the
application.
GOOD - OK 0x80 No error has been diagnosed.
GOOD - Maintenance
demanded
0xA8 The measured value is valid.
It is highly advisable to service the device in the near
future.
GOOD - Function check 0xBC The measured value is valid.
The measuring device is performing an internal
function check. The function check does not have any
noticeable effect on the process.
9.3.4 Factory setting
The slots are already assigned in the automation system for initial commissioning.
Assigned slots
Slot Factory setting
1 Mass flow
2 Volume flow
3 Corrected volume flow
4 Density
5 Reference density