Technical data Proline Promass I 300 PROFINET
230 Endress+Hauser
16.3 Input
Measured variable Direct measured variables
Mass flow
Density
Temperature
Viscosity
Calculated measured variables
Volume flow
Corrected volume flow
Reference density
Measuring range Measuring range for liquids
DN Measuring range full scale values 
min(F)
to 
max(F)
[mm] [in] [kg/h] [lb/min]
8 ³⁄₈ 0 to 2 000 0 to 73.50
15 ½ 0 to 6 500 0 to 238.9
15 FB ½ FB 0 to 18 000 0 to 661.5
25 1 0 to 18 000 0 to 661.5
25 FB 1 FB 0 to 45 000 0 to 1 654
40 0 to 45 000 0 to 1 654
40 FB 1½ FB 0 to 70 000 0 to 2 573
50 2 0 to 70 000 0 to 2 573
50 FB 2 FB 0 to 180 000 0 to 6 615
80 3 0 to 180 000 0 to 6 615
FB = Full bore
Measuring range for gases
The full scale value depends on the density and the sound velocity of the gas used and can
be calculated with the formula below:
max(G)
= minimum (
max(F)
· ρ
G
: x ; ρ
G
· c
G
· π/2 · (d
i
)
2
· 3600)
max(G)
Maximum full scale value for gas [kg/h]
max(F)
Maximum full scale value for liquid [kg/h]
max(G)
< 
max(F)
max(G)
can never be greater than 
max(F)
ρ
G
Gas density in [kg/m³] at operating conditions
x Constant dependent on nominal diameter
c
G
Sound velocity (gas) [m/s]
d
i
Measuring tube internal diameter [m]