Proline Promass I 300 PROFINET Technical data
Endress+Hauser 237
Light emitting diodes (LED)
Status information Status indicated by various light emitting diodes
The following information is displayed depending on the device version:
Supply voltage active
Data transmission active
Device alarm/error has occurred
PROFINET network available
PROFINET connection established
PROFINET blinking feature
Diagnostic information via light emitting diodes →  155
Low flow cut off The switch points for low flow cut off are user-selectable.
Galvanic isolation The outputs are galvanically isolated from one another and from earth (PE).
Protocol-specific data
Protocol Application layer protocol for decentral device periphery and distributed
automation, Version 2.3
Communication type 100 MBit/s
Conformity class Conformance Class B
Netload Class Netload Class II
Baud rates Automatic 100 Mbit/s with full-duplex detection
Cycle times From 8 ms
Polarity Auto-polarity for automatic correction of crossed TxD and RxD pairs
Media Redundancy Protocol
(MRP)
Yes
System redundancy support System redundancy S2 (2 AR with 1 NAP)
Device profile Application interface identifier 0xF600
Generic device
Manufacturer ID 0x11
Device type ID 0x843B
Device description files (GSD,
DTM, DD)
Information and files under:
www.endress.com
On the product page for the device: Documents/Software → Device drivers
www.profibus.org
Supported connections 2 x AR (IO Controller AR)
1 x AR (IO-Supervisor Device AR connection allowed)
1 x Input CR (Communication Relation)
1 x Output CR (Communication Relation)
1 x Alarm CR (Communication Relation)
Configuration options for
measuring device
DIP switches on the electronics module, for device name assignment (last
part)
Manufacturer-specific software (FieldCare, DeviceCare)
Web browser
Device master file (GSD), can be read out via the integrated Web server of
the measuring device
Configuration of the
device name
DIP switches on the electronics module, for device name assignment (last
part)
DCP protocol
Process Device Manager (PDM)
Integrated Web server