Proline Promass I 300 PROFINET Operation options
Endress+Hauser 71
213
4
3
4
5 6 7
A0034570
1 Transmitter with integrated WLAN antenna
2 Transmitter with external WLAN antenna
3 LED lit constantly: WLAN reception is enabled on measuring device
4 LED flashing: WLAN connection established between operating unit and measuring device
5 Computer with WLAN interface and Web browser (e.g. Microsoft Internet Explorer, Microsoft Edge) for
accessing the integrated device Web server or with operating tool (e.g. FieldCare, DeviceCare)
6 Mobile handheld terminal with WLAN interface and Web browser (e.g. Microsoft Internet Explorer, Microsoft
Edge) for accessing the integrated device Web server or operating tool (e.g. FieldCare, DeviceCare)
7 Smart phone or tablet (e.g. Field Xpert SMT70)
Function WLAN: IEEE 802.11 b/g (2.4 GHz)
Encryption WPA2-PSK AES-128 (in accordance with IEEE 802.11i)
Configurable WLAN channels 1 to 11
Degree of protection IP67
Available antennas Internal antenna
External antenna (optional)
In the event of poor transmission/reception conditions at the place of
installation.
Only one antenna active in each case!
Range Internal antenna: typically 10 m (32 ft)
External antenna: typically 50 m (164 ft)
Materials (external antenna) Antenna: ASA plastic (acrylic ester-styrene-acrylonitrile) and nickel-
plated brass
Adapter: Stainless steel and nickel-plated brass
Cable: Polyethylene
Connector: Nickel-plated brass
Angle bracket: Stainless steel
Configuring the Internet protocol of the mobile terminal
NOTICE
If the WLAN connection is lost during the configuration, settings made may be lost.
Make sure that the WLAN connection is not disconnected while configuring the device.
NOTICE
In principle, avoid simultaneous access to the measuring device via the service
interface (CDI-RJ45) and the WLAN interface from the same mobile terminal. This
could cause a network conflict.
Only activate one service interface (CDI-RJ45 service interface or WLAN interface).
If simultaneous communication is necessary: configure different IP address ranges, e.g.
192.168.0.1 (WLAN interface) and 192.168.1.212 (CDI-RJ45 service interface).