EasyManua.ls Logo

Endress+Hauser Proline Prosonic Flow P 500 - Load; Ex Connection Data

Endress+Hauser Proline Prosonic Flow P 500
88 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Proline Prosonic Flow P 500
Endress+Hauser 27
Load
Output signal →  21
Ex connection data Safety-related values
Order code
"Output; input 1"
Output type Safety-related values
"Output; input 1"
26 (+) 27 (–)
Option BA Current output
4 to 20 mA HART
U
N
= 30 V
DC
U
M
= 250 V
AC
Option MA Modbus RS485 U
N
= 30 V
DC
U
M
= 250 V
AC
Order code
"Output; input 2";
"Output; input 3"
Output type Safety-related values
Output; input 2 Output; input 3
24 (+) 25 (–) 22 (+) 23 (–)
Option B Current output
4 to 20 mA
U
N
= 30 V
DC
U
M
= 250 V
AC
Option D User-configurable input/
output
U
N
= 30 V
DC
U
M
= 250 V
AC
Option E Pulse/frequency/switch
output
U
N
= 30 V
DC
U
M
= 250 V
AC
Option F Pulse output, phase-
shifted
U
N
= 30 V
DC
U
M
= 250 V
AC
Option H Relay output U
N
= 30 V
DC
I
N
=100 mA
DC
/500 mA
AC
U
M
= 250 V
AC
Option I Current input 4 to 20 mA U
N
= 30 V
DC
U
M
= 250 V
AC
Option J Status input U
N
= 30 V
DC
U
M
= 250 V
AC
Intrinsically safe values
Order code
"Output; input 1"
Output type Intrinsically safe values
"Output; input 1"
26 (+) 27 (–)
Option CA Current output
4 to 20 mA HART Ex i
passive
U
i
= 30 V
l
i
= 100 mA
P
i
= 1.25 W
L
i
= 0 µH
C
i
= 6 nF
Option CC Current output
4 to 20 mA HART Ex i
active
Ex ia
1)
U
0
= 21.8 V
l
0
= 90 mA
P
0
= 491 mW
L
0
= 4.1 mH (IIC)/15 mH
(IIB)
C
0
= 160 nF (IIC)/
1 160 nF (IIB)
U
i
= 30 V
l
i
= 10 mA
P
i
= 0.3 W
L
i
= 5 µH
C
i
= 6 nF
1) Only available for Proline 500 transmitter Zone 1; Class I, Division 1.

Table of Contents