Measuring system
The measuring system consists of a transmitter and two or one sensor sets. The transmitter and
sensor sets are mounted in physically separate locations. They are interconnected by sensor cables.
The measuring system uses a measurement method based on the transit time difference. Here, the
sensors function as sound generators and sound receivers. Depending on the application and version,
the sensors can be arranged for a measurement via 1, 2, 3 or 4 traverses → 8.
The transmitter serves to control the sensor sets, to prepare, process and evaluate the measuring
signals, and to convert the signals to the desired output variable.
Transmitter
1 Transmitter with integrated ISEM
2 Sensor cables
• Electronics and ISEM (intelligent sensor electronics module) in the transmitter housing
• Signal transmission: analog
Order code for "Integrated ISEM electronics", option B: transmitter
Sensor cables
Sensor cables can be ordered in various lengths → 83
• Length: max. 30 m (90 ft)
• Cable with a common shield and individual shielded cores
Ex Zone
Use in: Ex Zone 1 and 2; Class 1, Division 2 and Class 1, Division 1
Housing versions and materials
• Transmitter housing
• Aluminum, coated: aluminum, AlSi10Mg, coated
• Cast, stainless: cast, stainless steel, 1.4409 (CF3M) corresponds to the properties of 316L
• Window material: glass
Configuration
• External operation via 4-line, illuminated graphic local display (LCD) with touch control and
guided menus ("Make-it-run" wizards) for application-specific commissioning.
• Via service interface or WLAN connection:
• Operating tools (e.g. FieldCare, DeviceCare)
• Web server (access via web browser)