Proline Prosonic Flow P 500
Endress+Hauser 59
Display modules
–40 to +80 °C (–40 to +176 °F)
Relative humidity
The device is suitable for use in outdoor and indoor areas with a relative humidity of 5 to 40%.
Operating height
According to EN 61010-1
• ≤ 2 000 m (6 562 ft)
• > 2 000 m (6 562 ft) with additional overvoltage protection (e.g. Endress+Hauser HAW Series)
Degree of protection Transmitter
• IP66/67, Type 4X enclosure, suitable for pollution degree 4
• When the housing is open: IP20, Type 1 enclosure, suitable for pollution degree 2
• Display module: IP20, Type 1 enclosure, suitable for pollution degree 2
Sensor
Order code for "Sensor version", options AA, AB, AC, AD, AE:
• IP68, Type 6P enclosure, suitable for pollution degree 4
• For the operation of the device under water
• Operating duration at a maximum depth of:
• 3 m (10 ft): permanent use
• 10 m (30 ft): maximum 48 hours
Order code for "Sensor version", options AG, AH:
IP66/67, Type 4X enclosure, suitable for pollution degree 4
When the housing is open: IP20, Type 1 enclosure, suitable for pollution degree 2
Optional
External WLAN antenna
IP67
Shock and vibration
resistance
Vibration sinusoidal, in accordance with IEC 60068-2-6
• 2 to 8.4 Hz, 7.5 mm peak
• 8.4 to 2 000 Hz, 2 g peak
Vibration broad-band random, according to IEC 60068-2-64
• 10 to 200 Hz, 0.01 g
2
/Hz
• 200 to 2 000 Hz, 0.003 g
2
/Hz
• Total: 2.70 g rms
Shock half-sine, according to IEC 60068-2-27
6 ms 50 g
Rough handling shocks according to IEC 60068-2-31
Electromagnetic
compatibility (EMC)
As per IEC/EN 61326 and NAMUR Recommendation 21 (NE 21) and 43 (NE43)
Details are provided in the Declaration of Conformity.
This unit is not intended for use in residential environments and cannot guarantee adequate
protection of the radio reception in such environments.
For detailed information on high-temperature sensors CH-050 / CH-100 (order code for
"Sensor version", options AG, AH), see Special Documentation "High-temperature" → 87.