Proline Prosonic Flow P 500
60 Endress+Hauser
Process
Medium temperature range
Sensor version Frequency Temperature
C-030-A 0.3 MHz –40 to +100 °C (–40 to +212 °F)
C-050-A 0.5 MHz –20 to +80 °C (–4 to +176 °F)
C-100-A 1 MHz –20 to +80 °C (–4 to +176 °F)
C-200-A 2 MHz –20 to +80 °C (–4 to +176 °F)
C-500-A 5 MHz –40 to +150 °C (–40 to +302 °F)
C-100-B 1 MHz –40 to +80 °C (–40 to +176 °F)
C-200-B 2 MHz –40 to +80 °C (–40 to +176 °F)
C-100-C 1 MHz 0 to +170 °C (+32 to +338 °F)
C-200-C 2 MHz 0 to +170 °C (+32 to +338 °F)
CH-050-A 0.5 MHz +150 to +220 °C (302 to +428 °F): Order code for "Process temperature", option H
+210 to +370 °C (410 to +698 °F): Order code for "Process temperature", option I
+350 to +550 °C (+662 to +1 022 °F): Order code for "Process temperature", option J
CH-100-A 1 MHz +150 to +220 °C (302 to +428 °F): Order code for "Process temperature", option H
+210 to +370 °C (410 to +698 °F): Order code for "Process temperature", option I
+350 to +550 °C (+662 to +1 022 °F): Order code for "Process temperature", option J
Sound velocity range
600 to 3 000 m/s (1 969 to 9 843 ft/s)
Medium pressure range
No pressure limitation For correct measurement, the static pressure of the medium must be higher
than the vapor pressure.
Pressure loss
There is no pressure loss.