Proline Prosonic Flow P 500
58 Endress+Hauser
Special mounting
instructions
Weather protection cover
146 (5.75)
48 (1.9)
12 (0.47)
280 (11.0) 255 (10.0)
134 (5.3)
30 (1.18)
A0029553
 56 Weather protection cover for Proline 500; engineering unit mm (in)
Environment
Ambient temperature range
Transmitter Standard:–40 to +60 °C (–40 to +140 °F)
Optional order code for "Test, certificate", option JN:
–50 to +60 °C (–58 to +140 °F)
Readability of the local display –20 to +60 °C (–4 to +140 °F)
The readability of the display may be impaired at temperatures outside
the temperature range.
Sensor DN 15 to 65 (½ to 2½")
–40 to +150 °C (–40 to +302 °F)
DN 50 to 4000 (2 to 160")
Standard:–40 to +80 °C (–40 to +176 °F)
Optional: 0 to +170 °C (+32 to +338 °F)
DN 50 to 600 (2 to 24")
High temperature: +150 to +550 °C (+302 to +1 022 °F)
Sensor cable (connection between
transmitter and sensor)
DN 15 to 65 (½ to 2½")
Standard (TPE
1)
): –40 to +80 °C (–40 to +176 °F)
DN 50 to 4000 (2 to 160")
Standard (TPE halogen-free): –40 to +80 °C (–40 to +176 °F)
Optional (PTFE
1)
): –50 to +170 °C (–58 to +338 °F)
1) Armored version also available for order
If operating outdoors:
Avoid direct sunlight, particularly in warm climatic regions.
In principle, it is permitted to insulate the sensors mounted on the pipe. In the case of insulated
sensors, make sure that the process temperature does not exceed or drop below the specified
cable temperature.
For notes on insulation for high-temperature sensors, see the Special Documentation on high-
temperature applications for the device →  87
You can order a weather protection cover from Endress+Hauser. →  83.
Storage temperature
The storage temperature for all components (except display modules and order code for "Sensor
version", options AG, AH) corresponds to the ambient temperature range→  58.
Order code for "Sensor version", options AG, AH: –50 to +80 °C (–58 to +176 °F)