EasyManua.ls Logo

Endress+Hauser Proline Prosonic Flow P 500 - Page 56

Endress+Hauser Proline Prosonic Flow P 500
88 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Proline Prosonic Flow P 500
56 Endress+Hauser
8. Insert the sensor cable into each individual sensor until the end stop and tighten the locking
nut.
A0043386
 53 Mounting the sensors and connecting the sensor cables
This completes the mounting procedure. The sensors can now be connected to the
transmitter via the sensor cables and the error message can be checked in the sensor
check function.
To ensure good acoustic contact, the visible measuring pipe surface must be clean and free
from flaking paint and/or rust.
If the sensor is removed from the measuring pipe, it must be cleaned and new coupling gel
applied (if there is no coupling pad).
On rough measuring pipe surfaces, the gaps in the rough surface must be filled with
sufficient amounts of coupling gel if use of the coupling pad does not suffice (installation
quality check).

Table of Contents

Other manuals for Endress+Hauser Proline Prosonic Flow P 500

Related product manuals