Proline Prosonic Flow P 500
56 Endress+Hauser
8. Insert the sensor cable into each individual sensor until the end stop and tighten the locking
nut.
A0043386
53 Mounting the sensors and connecting the sensor cables
This completes the mounting procedure. The sensors can now be connected to the
transmitter via the sensor cables and the error message can be checked in the sensor
check function.
• To ensure good acoustic contact, the visible measuring pipe surface must be clean and free
from flaking paint and/or rust.
• If the sensor is removed from the measuring pipe, it must be cleaned and new coupling gel
applied (if there is no coupling pad).
• On rough measuring pipe surfaces, the gaps in the rough surface must be filled with
sufficient amounts of coupling gel if use of the coupling pad does not suffice (installation
quality check).