EasyManua.ls Logo

Endress+Hauser Proline Prosonic Flow P 500 - Page 77

Endress+Hauser Proline Prosonic Flow P 500
88 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Proline Prosonic Flow P 500
Endress+Hauser 77
Supported operating
tools
Operating unit Interface Additional information
Web browser Notebook, PC or tablet
with web browser
CDI-RJ45 service
interface
WLAN interface
Special Documentation for
device
DeviceCare SFE100 Notebook, PC or tablet
with Microsoft Windows
system
CDI-RJ45 service
interface
WLAN interface
Fieldbus protocol
→  85
FieldCare SFE500 Notebook, PC or tablet
with Microsoft Windows
system
CDI-RJ45 service
interface
WLAN interface
Fieldbus protocol
→  85
Field Xpert SMT70/77/50 All Fieldbus protocols
WLAN interface
Bluetooth
CDI-RJ45 service
interface
Operating Instructions
BA01202S
Device description files:
Use update function of
handheld terminal
SmartBlue App Smart phone or tablet
with iOs or Android
WLAN →  85
Other operating tools based on FDT technology with a device driver such as DTM/iDTM or
DD/EDD can be used for device operation. These operating tools are available from the
individual manufacturers. Integration into the following operating tools, among others, is
supported:
FactoryTalk AssetCentre (FTAC) from Rockwell Automation → www.rockwellautomation.com
Process Device Manager (PDM) from Siemens → www.siemens.com
Asset Management Solutions (AMS) from Emerson → www.emersonprocess.com
FieldCommunicator 375/475 from Emerson → www.emersonprocess.com
Field Device Manager (FDM) from Honeywell → www.process.honeywell.com
FieldMate from Yokogawa → www.yokogawa.com
PACTWare → www.pactware.com
The related device description files are available: www.endress.com → Download Area
Web server
With the integrated web server, the device can be operated and configured via a web browser service
interface (CDI-RJ45) or WLAN interface. The structure of the operating menu is the same as for the
local display. In addition to the measured values, status information on the device is displayed and
can be used to monitor device health. Furthermore the device data can be managed and the network
parameters can be configured.
A device that has a WLAN interface (can be ordered as an option) is required for the WLAN
connection: order code for "Display; operation", option G "4-line, illuminated; touch control + WLAN".
The device acts as an Access Point and enables communication by computer or a mobile handheld
terminal.
Supported functions
Data exchange between the operating unit (such as a notebook, for example,) and measuring device:
Upload the configuration from the measuring device (XML format, configuration backup).
Save the configuration to the measuring device (XML format, restore configuration).
Export event list (.csv file)
Export parameter settings (.csv file or PDF file, document the measuring point configuration)
Export the Heartbeat verification report (PDF file, only available with the Heartbeat Verification
→  82 application package)
Flash firmware version for device firmware upgrade, for example
Download driver for system integration
Visualize up to 1000 saved measured values (only available with the Extended HistoROM
application package→  82)

Table of Contents

Other manuals for Endress+Hauser Proline Prosonic Flow P 500

Related product manuals