Solicap M FTI55, FTI56 Technical data
Endress+Hauser 87
TI418F66
Fully insulated (PA):
Process pressure limits
–1 to 25 bar (observe dependencies => process connections from Page 18 ff. and operating
conditions: process from Page 84.)
The lowest value from the derating curves of the device and the selected flange applies.
Please refer to the following standards for the pressure values permitted at higher temperatures:
pR EN 1092-1: 2005 Table, Appendix G2
With regard to its resistance/temperature property, the material 1.4435 is identical to 1.4404
(AISI 316L) which is grouped under 13E0 in EN 1092-1 Tab. 18. The chemical makeup of the
two materials can be identical.
ASME B 16.5a- 1998 Tab. 2-2.2 F316
ASME B 16.5a- 1998 Tab. 2.3.8 N10276
JIS B 2220
10.5.3 Temperature-derating separate housing
BA300Fxx021
T
a
: Ambient temperature
T
P
: Process temperature
* temperature at the separate housing 70 °C
barbar
PP
pp
00
–40–40 –20–20–60–60
8080 120120
4040
2020
2020 4040 6060 100100 140140
TT
PP
°C°C
00
T
a
T
P
0
0
T
P
–40 –20–60
80 120 160
80
60
40
20
–80 200
–60
–40
–20
20 40 60 100 140
+180/
+110
+170
+120
/
°C
T
a
120
100
°C
*
180