48
1600751-4D
11.2.5.a.1 Rheobase/ Chronaxie
Select Rheobase /
Chronaxie with the
central controller [5]
and enter with the
confirmation button [6]
Start with the
Rheobase. The phase
duration is fixed on
500ms.
To start push the
confirmation button [6]
you activate the
intensity setting. Use
the central controller [5]
to adjust the intensity.
Confirm the value by
pushing the
confirmation button [6]
Repeat by pushing the
confirmation button [6]
again.
Select Chronaxie with
the central controller [5]
and confirm with the
confirmation button [6]
The retrieved intensity
in the Rheobase will be
doubled automatic.
Adjust the Phase
duration with the
central controller [5]
until the muscle
contract.
Electrotherapy
2
Diagnostics
Electrode selection
Phase Interval
5 s
Rheobase / Chronaxie
S/D Curve Rectangular
S/D Curve Triangular
Report
Beep
mA
Diagnostics
2
Rheobase / Chronaxie
Rheobase
500 ms
Chronaxie
5.0 mA
Diagnostics
2
Rheobase / Chronaxie
Rheobase
500 ms
Chronaxie
1000 µs
10.0 mA

Table of Contents

Related product manuals