45
Setting
Range
Limitations
Operating mode
AUTO / MANUAL
In manual mode the insufflation
and exsufflation are triggered by
the user.
Inh. Pressure
(cmH2O)
5
-
70
None
Slope
0
-
5
None
Inh Time
(s)
0.5
-
5
Only applies in
AUTO
mode.
Insp. Oscillation Freq.
(Hz)
4
-
20
None
Insp. Oscillation Amp.
1
-
3
None
Trigger
OFF/1
-
3
Only applies in AUTO mode.
Not
compatible with timed Pause.
Exh. Pressure
(cmH2O)
0 to
-
70
None
Exh. Time
(s)
0.5
-
5
Only
applies in AUTO mo
de.
Exhal. Oscillation Freq.
(Hz)
4
-
20
None
Exhal. Oscillation Amp.
1
-
3
None
Pause
(s)
OFF / 0.5
-
5
Only applies in AUTO mode.
No
t
compatible with Trigger. (Hidden
when Trigger is set)
PEEP
(cmH2O)
OFF / 1 to 20
None
Cycles Nb
1 to 20
Only applies in AUTO mode.
Treatment End
Inhal
ation/Exhalation
Only applies in A
UTO
mode
IPPB
The IPPB Mode delivers the therapy accordingly to the settings of: Inhalation Flow, Pressure Max,
Inhalation Max Time, Inhalation Slope, PEEP, Exhalation Fall and Inhalation Trigger.
The therapy delivered will follow the sequence:
Inhalation flow during inhalation
Exhalation Fall and PEEP during exhalation
During the inhalation phase, the device delivers the Inhalation Flow until the device reaches one of
the two limits: Pressure Max or Inhalation Max Time.
When Inhalation Slope is ON, the device will decrease the Inhalation Flow as pressure is increasing in
order to reach the Pressure Max when the flow decreased at half of the Inhalation Flow.
During the exhalation phase, the device switch to the set PEEP decreasing the pressure accordingly
with the Exhalation Fall.
The Trigger function allows the patient to start the therapy during the exhalation. The touch pad can
be used to trigger the inhalation.
The treatment time, if activated, will automatically stop the treatment at the end of the set time.
Setting
Range
Limitations
Inh. Flow
(l/min)
5
-
100
None
Inh. Slope
ON/OFF
None
Pressure Max.
(cmH2O)
10
-
50
None
Inh. Trig.
OFF / 1
-
5
None
PEEP
(cmH2O)
OFF / 1
-
20
None
Exh Slope
0
-
5
None
Inh. Max Time (s)
0.5
-
20
None
Treatment
time (min)
OFF/
5
-
30
None