11 / 66
2 Safety Instructions
80110-801_V23493
2022-03
Risks due to the environment
NOTICE
Unsuitable temperature or level of humidity during operation
If you operate the unit at an unsuitable temperature or level of hu-
midity, it may sustain damage, fail, or not perform properly.
Operate the unit at a suitable temperature and level of humidity.
You will find the tolerances for temperature and humidity in the
Technical Data.
If other ambient conditions must be observed for operation of the
unit, you will also find them in the Technical Data.
NOTICE
Unsuitable temperature or humidity in transit or storage
If you transport or store the unit at an unsuitable temperature or level
of humidity, it may sustain damage and fail.
Transport and store the unit at a suitable temperature and level
of humidity. You will find the tolerances for temperature and hu-
midity in the Technical Data.
If other ambient conditions must be observed for transport and
storage of the unit, you will also find them in the Technical Data.
NOTICE
Insufficient acclimatization time, unsuitable temperature during
acclimatization
If the unit was stored or transported below or above a certain tem-
perature, it will take a certain time and temperature to acclimatize.
If you do not observe the rules, the unit can sustain damage and fail.
Acclimatize the unit according to the rules in the Technical Data.
NOTICE
Overheating of the unit due to poor ventilation
If ventilation is poor, the unit can overheat, sustain damage, and fail.
Install the unit in such a way that there is an unobstructed circu-
lation of air around the housing. Installation in confined wall re-
cesses is prohibited.
WARNING
Penetration of fluid into the unit during operation or when clean-
ing the unit
Risk of electric shock to the medical personnel.
The unit may become damaged and fail.
Make sure no liquid can penetrate the unit.
Do not place vessels containing liquids on top of the unit.
Take care when connecting the separating medium to the pump
unit that the unit does not come into contact with the separating
medium. The same applies when removing the separating medi-
um.