2 • Safety Instructions
80110-801_V23493
2022-03
Activation of the unit with no knowledge of active settings
If the user does not understand the active settings of the unit, he can
cause the patient accidental tissue damage.
Check the active settings on the unit display after setting the unit,
changing a program, or changing a setting.
Uncontrolled water jet
Depending on the effect setting, the water jet can damage skin or the
eyes.
Never point the water jet at medical personnel. Do not point the
water jet at the patient accidentally.
Wear goggles for your own safety.
Point the tip of the applicator at the operating field. Only press the
pedal when you want to cut.
During the priming process, separating medium can escape from
the applicator.
Depending on the effect setting, the water jet can damage skin or the
eyes.
Never point the applicator at medical personnel. Never point the
applicator at the patient during the priming process.
Wear goggles for your own safety.
If separating medium escapes from the applicator, stop the prim-
ing process. Press the Selection button next to the menu item.
STOP.
Applicator pressure hose kinked
If you kink the pressure hose you will obtain an error message in crit-
ical cases and have to replace the applicator and pump unit.
Do not kink the pressure hose.
Notes
Grounding
Note: If necessary, connect the grounding pin of the unit or the cart to the grounding
system of the operating room using a grounding cable.
Use of a defibrillator
Note: The unit conforms to the requirements of Type CF and is protected against the
effects of a defibrillator discharge.
Filled pump cartridge
Note: Do not slide a filled pump cartridge into the pump shaft.
Selecting effect and suction
capacity
Note: Select the effect and the suction capacity according to the application, the tissue
properties and the volumes of fluid to be delivered.
Apply settings individually for the patient, taking into account the conditions specified
above.