6 Installation
80110-801_V23493
2022-03
WARNING
Damaged unit, damaged accessories, modified unit, and modified
accessories
Risk of electric shock to the patient and medical personnel! Risk of
damage to property.
Risk of injury to the patient and medical personnel! Risk of damage
to property.
Check the unit and accessories (e.g. footswitch, cables, tubes from
instruments) for damage every time before use.
You must not use a damaged unit or damaged accessories. Re-
place defective accessories.
If the unit or accessories are damaged, please contact our cus-
tomer service.
Blown power fuses may only be replaced by a competent techni-
cian.
For your safety and that of the patient: Never attempt to perform
repairs or make modifications yourself. Any modification will in-
validate liability on the part of Erbe Elektromedizin GmbH.
Installing ERBEJET 2
To ensure the unit stands safely and free of vibration, place it on a stable, hori-
zontal, vibration-free surface.
Avoid direct sunlight for reasons of heating and good display legibility.
Turn the rear of the unit away from the operating field.
Lock the wheel brakes on the front wheels.
The supply voltage must match the voltage specified on the unit's rating plate.
Connect the unit to a properly installed, grounded power outlet. Use the compati-
ble Erbe power cord (see the Accessories, Compatible power cords chapter). If the
unit is installed on the VIO CART, connect to the power supply using the VIO
CART’s power cord.
Install the unit such that the power cord could be easily disconnected from the
power source.