Status Messages, Error Messages
Status messages
Status messages are of an informative nature or prompt you to take operational ac-
tion.
Error messages
An error message consists of an error code (EC.) and an error text. The ERBEJET 2 dis-
play shows two different types of error messages.
a) Error messages that prompt you to take action and remedy the error. You will find
these error messages in the table.
b) Error messages that prompt you to inform Technical Service. These error messages
are not described individually in the user manual. The error text reads:
• Please inform Technical Service.
Status message Description
Suction Module recognized! You have switched on the ESM 2 suction module at the power switch or have
connected with the ECB cable.
Ready The pump unit is connected. ERBEJET 2 is ready for activation.
The footswitch is ready for operation. You have connected the footswitch.
The footswitch has been disconnected. You have removed the footswitch.
ERBEJET has cooled down. / Activation is
possible again!
ERBEJET 2 has cooled down sufficiently after overheating so that activation is
possible again.
Press "START" to prime. Press the Selection button next to the menu item START to prime the pump
unit with separating medium.
Priming running Priming running
Priming completed. Priming completed.
Priming aborted You have aborted the priming by pressing the STOP button.
Remove used cartridge! Remove the pump unit from the pump shaft.
Please insert new cartridge! Slide the unused pump unit into the pump shaft.
Connecting cartridge - Please wait. Wait until the pump unit is disconnected.
Uncoupling cartridge - Please wait. Wait until the pump unit is unlocked.
Referencing running - Please wait. The pump unit drive can be in various positions after switching on ERBEJET 2.
The drives therefore start by moving right to the back (reference position).
Referencing completed. The pump unit drives have reached the reference position.