5 Status Messages, Error Messages
40 / 66
80110-801_V23493
2022-03
Please note the error code. Give the error code to Technical Service.
You will find error messages with error code outputs in numerical order.
Error messages with error codes in numerical order
Error messages
EC. Error text Possible cause(s) What should you do?
J-5D ESM 2: Software incompati-
ble!
The ESM 2 suction module has non-com-
patible software.
Use an ESM 2 suction module with
compatible software. Inform Techni
-
cal Service.
J-5E FS: Software incompatible! The footswitch has incompatible software. Use a footswitch with compatible
software. Inform Technical Service.
J-06 Activation has been inter-
rupted.
System error Reactivate ERBEJET 2. If the error
occurs frequently, inform Technical
Service.
J-07 Please check the position of
the cartridge in the slot!
The pump unit may have been pushed out
of the pump shaft.
If the pump unit is not correctly
seated in the pump shaft, firstly
attempt to press the pump unit back
in place. The error message disap
-
pears if successful. If this action
fails, press the selection button next
to the EJECT display. Remove the
pump unit and applicator. Take a new
pump unit and a new applicator. If
the error occurs frequently, inform
Technical Service.
J-20 Module conn. to ECB not
compatible.
A non-compatible unit is connected to the
ECB socket of ERBEJET 2.
Remove the unit by detaching the
ECB cable between the unit and
ERBEJET 2.
J-21 Check date/clock time
adjustment.
The battery is discharged. Date and time
may have been lost.
You can still work with ERBEJET 2,
but inform Technical Service none
-
theless.
J-23 Check date/clock time
adjustment.
Date and time are inconsistent. You can still work with ERBEJET 2,
but inform Technical Service none-
theless.
J-24 Module conn. to ECB not
compatible.
A non-compatible footswitch is connected
to ERBEJET 2.
Remove the footswitch.
Error text Possible cause(s) What should you do?
Suction: activation inter-
rupted.
System error Reactivate the ESM 2 suction module. If
the error occurs frequently, inform Tech
-
nical Service.
Suction OFF You have attempted to activate the ESM 2 suction
module, but the suction is deactivated in the Suc-
tion window of the ERBEJET 2 display. Suction
mode: OFF
Set the suction mode to PEDAL.