5 • Status Messages, Error Messages
80110-801_V23493
2022-03
Suction not ready for oper-
ation!
Various causes are possible:
a) The ESM 2 suction module is not connected.
b) There is no connection between ERBEJET 2
and the ESM 2 suction module.
c) System error.
a) Switch on the ESM 2 suction module
with the power switch.
b) Check the connection between ERBE-
JET 2 and the ESM 2 suction module
(ECB cable).
c) Inform Technical Service.
Please interrupt activation. a) You keep the pedal depressed. b) System error a) Release the pedal. b) Inform Technical
Service.
ERBEJET 2 not ready for
operation!
a) You have attempted to activate ERBEJET 2
before the pump unit drives were ready.
b) You have attempted to activate ERBEJET 2
without a pump unit.
a) Do not activate until the following
message is displayed: "Ready."
b) Insert a pump unit.
ERBEJET overheated /
Activation not possible!
The temperature of the pump unit drives is too
high. The drives have been switched off.
Let ERBEJET 2 cool down until the fol-
lowing message is displayed:
"ERBEJET has cooled down! / Activation
is possible again!"
Error: Powerfail System error Reactivate ERBEJET 2. If the error occurs
frequently, inform Technical Service.
The footswitch is not ready. Various causes are possible:
a) No footswitch is connected.
b) The footswitch has incompatible software.
a) Connect a footswitch to ERBEJET 2.
b) Use a footswitch with compatible soft-
ware.
c) Inform Technical Service.
Instrument damaged or
pump reused. Replace
pump and instrument.
Various causes are possible:
a) The applicator nozzle is clogged.
b) You are attempting to connect a used pump
unit to the drive.
c) The drive is blocked or too stiff.
d) The drive mechanism is defective.
Press the selection button next to the
EJECT display. Remove the pump unit
and applicator. Take a new pump unit and
a new applicator. If the action fails,
inform Technical Service.
Critical temperature
reached! / Let ERBEJET
cool down!
The temperature of the pump unit drives is high.
ERBEJET 2 is ready for reactivation.
If possible, take activation breaks. Check
whether air can freely circulate around
the ERBEJET 2 housing.
You have exceeded the
max. ON time!
ON time limit attained. Options:
a) Observe the max ON time.
b) Extend the max ON time.
c) Switch off the ON time limit.
Error text Possible cause(s) What should you do?