4 • Working with ERBEJET 2
80110-801_V23493
2022-03
Chapter 4
Working with ERBEJET 2
Checking the unit and accessories
Damaged unit, damaged accessories, modified unit, and modified
accessories
Risk of electric shock to the patient and medical personnel! Risk of
damage to property.
Risk of injury to the patient and medical personnel! Risk of damage
to property.
Check the unit and accessories (e.g. footswitch, cables, tubes from
instruments) for damage every time before use.
You must not use a damaged unit or damaged accessories. Re-
place defective accessories.
If the unit or accessories are damaged, please contact our cus-
tomer service.
Blown power fuses may only be replaced by a competent techni-
cian.
For your safety and that of the patient: Never attempt to perform
repairs or make modifications yourself. Any modification will in-
validate liability on the part of Erbe Elektromedizin GmbH.
Connecting suction module ESM 2, connecting the applicator to the
pump unit, replacing the applicator
If you use an applicator with suction hose you can use the suction module ESM 2 for
suction.
Attaching the mounting bracket
Fig. 4-1
Equip the bracket with a mounting bracket.
Hanging the suction container
Note: The suction container must be positioned lower than the operating field so that
no fluid can flow back from the suction hose into the operating field.