4 Working with ERBEJET 2
28 / 66
80110-801_V23493
2022-03
1. Prime the pump unit and applicator with separating medium. Press the Selection
button next to the menu item. START. If you press the Selection button next to the
menu item SKIP, you skip over the priming process. You must then prime the
pump unit and applicator with separating medium by activating the water jet.
2. The menu item STOP must be activated if separating medium escapes from the
applicator during the priming process (see Safety instructions). If you activate the
menu item STOP before the priming process is completed, you must prime the
pump unit and applicator with separating medium by activating the water jet.
Mounting suction hose of applicator to suction container
Fig. 4-16
Note: The ESM 2 may be operated only in combination with a suction hose that is in-
tegrated in an applicator of the ERBEJET 2.
During the priming process, separating medium can escape from
the applicator.
Depending on the effect setting, the water jet can damage skin or
the eyes.
Never point the applicator at medical personnel. Never point the
applicator at the patient during the priming process.
Wear goggles for your own safety.
If separating medium escapes from the applicator, stop the
priming process. Press the Selection button next to the menu
item. STOP.
Tension on the suction hose
If you subject the suction hose to tension, secretion can escape
from the branch (1).
Do not subject the suction hose to tension.
1