2 • Safety Instructions
14 / 66
80110-801_V23493
2022-03
WARNING
Non-sterile infusion set
The patient may become infected if a non-sterile infusion set is used.
Use aseptic techniques when connecting the infusion set to a bot-
tle or a bag of separating medium.
Use aseptic techniques when connecting the infusion set to the
pump cartridge.
WARNING
A pump unit is used for several patients
Risk of cross-infection.
A pump unit may only be used for one patient.
WARNING
Non-sterile separating medium
The patient can be infected by the use of a non-sterile separating
medium.
Only use sterile, physiological saline solutions.
WARNING
Contaminated separating medium
Separating medium splash over the medical staff.
Risk of infection for the medical staff.
Avoid the accumulation of separating medium in the operating ar-
ea. Ensure adequate suction.
Use personal protective equipment, e.g. a face mask and splash
guard. The protective equipment must be appropriate for the risk
arising from the patient.
WARNING
Contaminated accessories
Risk of infection for the medical staff.
Wear gloves when dismantling the accessories
WARNING
Alternate use of disinfectant solutions based on different active
ingredients
The agents may affect one another. The disinfection effect of the dis-
infectant solution may be weakened. Danger of infection to the pa-
tient and medical personnel.
The housing plastic may become brittle and break. A color reaction
may occur with plastics.
Do not use these substances alternately.