18
5-6 Error message “Printer Cutter Error appears
Main causes when error message Printer Cutter Error appears:
A printer paper is stuck in cutter portion
Failure of thermal printer
Failure of printer control board assy
Are the printer papers
stuck in the printer portion?
YES
NO
Printer Cutter Error appears.
YES
OK
Does Printer Cutter Error still appear?
Remove the paper.
NO
Replace the thermal printer.
Replace the printer control board assy.