22
5-10 Error message "Right Lens 4-5 Error" or "Left Lens 4-5 Error"
appears
Main causes when error message "Right Lens 4-5 Error" or "Left Lens 4-5 Error"
appears:
Failure of LD-R assy or LD-L assy:
Note: If the error is not displayed frequently, the device can be used without replacing the LD-R
assy or the LD-L assy.
In such case, however, use the device after confirming that the error does not appear
anymore by pressing the clear key and performing the initial operation.
Right Lens 4-5 Error or
"Left Lens4-5 Error" appears.
YES
NO
OK
Replace the LD-R assy or
the LD-L assy.
Does "Right Lens 4-5 Error" or "Left Lens 4-5
Error" still appear even after pressing the
clear key and performing the initial operation?