31
5-18 Measurement result is not printed out/ printout is not normal
Main causes when measurement result is not printed out/ printout is not normal:
Failure of thermal printer
Failure of printer control board assy
Failure of control board assy
Replace the control board assy.
Replace the printer
control board assy.
YES
NO
OK
No print out/ printout is not normal.
Replace the thermal printer.
Is the printer printing normal?
Is the printer printing normal?
NO