23
5-11 Error message "Right Lens 6 In Error" or "Left Lens 6 In Error"
appears
Main causes when error message "Right Lens 6 In Error" or "Left Lens 6 In Error"
appears:
Failure of LD-R assy or LD-L assy
Note: If the error is not displayed frequently, the device can be used without replacing the LD-R
assy or the LD-L assy.
In such case, however, use the device after confirming that the error does not appear
anymore by pressing the clear key and performing the initial operation.
Right Lens6 In Error or
"Left Lens6 In Error" appears.
YES
OK
Does "Right Lens6 In Error" or "Left Lens6 In
Error" still appear even after pressing the
clear key and performing the initial operation.
NO
Replace the LD-R assy
or the LD-L assy.