5-2 CALIBRATION SETTING
Please allow the indicator to warm up at least 15 to 30 minutes before calibration.
Switch the capacity calibration switch to ON and the display shows .
Press the key and the display shows . then enter the setting mode.
5-2-1 Calibration Procedure
Zero Calibration
Weight Calibration
5-2-2 Zero Calibration
a) Make sure that there are no objects on the weighing platform and press
the key, after the indicator has stabilized the display will show
……. . The calibration is completed after about 5 seconds.
b) To skip the Zero calibration procedure, press the key.
5-2-3 Weight Calibration
a) Place an accurate weight onto the platform; use the front panel keys to key in
the weight value. Press the key, after the indicator is stabilized the display will
show ……. . The calibration is completed after about 5 seconds.
b) To skip the weight calibration, press the .
Zero Calibration
Complete
d
Confirmation
completed
Switch the calibration
switch to OFF
Place
calibrated
weight
Increase the number of
the flashing digit
Decrease the number
of the flashing digit
Shift the flashing digit
one space to left
Shift the flashing digit
one space to right
Save the settings
Exit the settings