EXCELL PRECISION CO., LTD
Setting value
Item Function
Parameter Description
Default
0 None
1 g
2 Kg
3 t
CSP-01 Unit
4 lb
2
0 None
1 1 Decimal Point
2 2 Decimal Point
CSP-02
Decimal
Point
3 3 Decimal Point
0
1
2
5
10
20
CSP-03 Division
50
Division size 1
CSP-04
Max.
Capacity
999999
↓
000000
Max. capacity 999999
CSP-05 Zero range
0 =full range
(±1%~30%)
Zero range = calibration zero point ± (Max.
capacity×setting value %)
0
CSP-06
Time of
zero
tracking
0.0 ~ 5.0
(sec)
Time and range of zero tracking should be
use at the same time. If the time is set to 0.0,
the zero tracking function is disabled
1.0
CSP-07
Range of
zero
tracking
0 ~ 9
Range of zero tracking = (setting value×½)D ,
D=min. division
Range and time of zero tracking should be
use at the same time. If the range is set to 0,
the zero tracking function is disabled
2
CSP-08
Investigate
time in stable
0.0 ~ 5.0
(sec)
Investigate time and range should be use at
the same time. If the time is set to 0.0, the
investigate time is disabled
1.0
CSP-09
Investigate
range in stable
0 ~ 9
Investigate time and range should be use at
the same time. If the range is set to 0, the
investigate range is disabled
2
0 Action
CSP-10
Weight
unstable,
function ZERO
and TARE
1 None
0
0 Action
CSP-11
Gross Weight
is negative,
function TARE
1 None
0