Chapter 6 Maintenance
6-1 Restore All Parameters to Their Default Factory Values.
(1) While the indicator is counting back to zero, adjust SW to ON and press keys.
(2) Display shows the flashing digits . .
(3) Confirm / abort.
(3-1) To confirm press the key & don’t release it until the display shows ,
then release the key and return the calibration SW to OFF.
(3-2) To abort, set the calibration SW to OFF directly.
6-2 Maintenance Function Parameters
Power on the machine. Press keys while the display counts back to zero.
.
Restore all parameters to their
default factory values. See for details
6-2-1
.
.
Clear zero compensation and tare
value. See for details 6-2-2.
.
Clear batch setting value. See
for details 6-2-3.
.
Display span voltage (mV/V)
See for details 6-2-5.
.
Display zero voltage (mV/V)
See for details 6-2-4.