EXCELL PRECISION CO., LTD
Chapter 2 General Function Guide
2-1 Function Setup and Operation Procedures
Function Operation Display Description
Enter calibration mode
Turn the calibration switch to “ON”
See 3-2 for details
Enter function setting
Press not release,
then press key
after the power is turned on
See 2-2 for details
Reset all parameters
back to default
Turn the power on then turn the
calibration switch to “ON” then
press and hold the
and keys during the
self-testing sequence
.
See 6-1 for details
Reset general function
parameters back to
default
Turn the power on and
press and keys
during self-testing sequence
See 6-2-1 for details
Clear zero point
compensation and tare
value
Turn the power on and
press and keys
during self-testing sequence, and
then press key
See 6-2-2 for details
Clear setpoint parameter
setting
Turn the power on and
press and keys
during self-testing sequence, and
then press two times
See 6-2-3 for details
Value of zero point
voltage(mV / V)
Turn the power on and
Press and ,then
press key three times
See 6-2-4 for details
Value of Span voltage
(mV / V)
Turn the power on and
Press and , then
Press key
See 6-2-5 for details
Entering to test mode
Turn the power on and
press and keys
during self-testing sequence
.
See 6-3 for details