37
04011210
EXCELL PRECISION CO., LTD
ZSME400000003
4-9-2 Hi, OK, Lo comparison
1. Normal HOLD ( FNC-11 = 0 )
Entering the Hold mode, Hi, OK, Lo comparison output. Escape Hold mode will switch
off the outputs.
2. Peak HOLD ( FNC-11 = 1, 2 )
If Peak Ready is ON, Hi, OK, Lo comparison output. Escape Hold mode will switch off
the outputs.
3. The absolute value of peak HOLD ( FNC-11 = 3 )
Entering the Hold mode, Hi, OK, Lo comparison output.
4. FNC-11 = 4
When the external input single Judgement is ON, Hi, OK, Lo comparison output.
5. If to release HOLD (display value and Hi, OK, Lo signal), please make HOLD release
signal on.
4-10 Totalizing (ACCU.) Auto / Transmit
With automatic totalising active (SQ-17) or RS232 / RS485 or BCD output
set to auto transmit.
1. SQ-01 = 1, 2, 4 or 5 batch / loss-in weight
a) When the weight reaches the Final weight and the batch finish signal is ON the
net weight will be added to the totaliser and number of additions is incremented.
The RS-232 / RS-485 and BCD outputs transmit data. (RS-01/ RS-02=10
accumulated weight and counts)
b) When the net weight returns to the zero range (FNC-10), then the sequence in a)
above can be repeated.
c) When SQ-01=1 or 4, Net Final and Batch finish=1, it accumulate once.
d) When SQ-01=2 or 5, -Net Final and Batch finish=1, it accumulate once.
2. SQ-01 = 3 Comparison mode
a) When the net weight exceeds the zero range and the weight has stabilized it will
be added to the totaliser and number of additions is incremented. The RS-232 –
RS-485 and BCD outputs transmit data.
3. SQ-01=6 HOLD mode
a).Lo. Wt. weightHi. Wt. weight will be added to the totaliser and number of
additions is incremented.RS-232 / RS-485 and BCD outputs transmit data. (RS-01/
RS-02=10 accumulated weight and counts)
b). Until the net weight returns to the zero range (FNC-10), then the sequence in a)
above can be repeated.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Excell EX2005 and is the answer not in the manual?

Excell EX2005 Specifications

General IconGeneral
BrandExcell
ModelEX2005
CategoryAccessories
LanguageEnglish

Related product manuals