EasyManua.ls Logo

FDE SO-CONNECT PID - Infusion Set

Default Icon
75 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Partie IV Syringe and infusion set
40
Connecting the syringe to the pump
To connect the SO
FILL syringe to the SO
CONNECT PAR
®
pump, please follow the steps below:
1. Turn the head of the pusher on the SO
CONNECT PAR
®
pump
2. Place the SO
FILL syringe on the SO
CONNECT PAR
®
pump and turn it clockwise.
Infusion set
Infusion sets with the characteristics below should be used with the SO
CONNECT PID pump and SO
FILL syringes:
small internal tube volume (optimum 0.1 ml, maximum 0.62 ml),
tube that does not exceed 90 cm length,
anti-throttle tube.
Warning
The use of an unsuitable infusion set may make the
SO
CONNECT PAR
®
pump malfunction, e.g. it may take longer to report an
occlusion.
Before using patient tubing accessories or connecting the
SO
CONNECT PAR
®
pump to other infusion systems, please contact
user support to identify the procedure to apply. Connecting the
SO
CONNECT PAR
®
pump to other infusion systems may make
the infusion rate inaccurate.

Table of Contents