Partie IV – Syringe and infusion set
40
Connecting the syringe to the pump
To connect the SO
♥
FILL syringe to the SO
♥
CONNECT PAR
®
pump, please follow the steps below:
1. Turn the head of the pusher on the SO
♥
CONNECT PAR
®
pump
2. Place the SO
♥
FILL syringe on the SO
♥
CONNECT PAR
®
pump and turn it clockwise.
Infusion set
Infusion sets with the characteristics below should be used with the SO
♥
CONNECT PID pump and SO
♥
FILL syringes:
▪ small internal tube volume (optimum 0.1 ml, maximum 0.62 ml),
▪ tube that does not exceed 90 cm length,
▪ anti-throttle tube.
Warning
▪
The use of an unsuitable infusion set may make the
SO
♥
CONNECT PAR
®
pump malfunction, e.g. it may take longer to report an
occlusion.
▪
Before using patient tubing accessories or connecting the
SO
♥
CONNECT PAR
®
pump to other infusion systems, please contact
user support to identify the procedure to apply. Connecting the
SO
♥
CONNECT PAR
®
pump to other infusion systems may make
the infusion rate inaccurate.