Titre VI – Alarmes
55
Alarm system
The table below summarises all the conditions in which the different alarms are generated.
Failure of a printed circuit board component detected
by the internal software of the SO
♥
CONNECT PID
pump
Motor blocked during infusion, due to excessively high
pressure in the tubing.
See section “Time required for
indicating an occlusion”
Syringe not detected during the infusion by the internal
software of the SO
♥
CONNECT PID pump with each
injection.
Remaining battery life less than 30 minutes.
The infusion will end in less than 5 minutes.
The infusion has been paused for more than 5 minutes
When two or more alarm conditions are met, the alarms are generated in the following order of priority:
▪ device failure,
▪ occlusion,
▪ syringe missing,
▪ end of infusion,
▪ battery low,
▪ infusion about to end,
▪ infusion in pause
High priority alarms are indicated by a message displayed on a flashing red background and a sound signal made up of 10 beeps every 10 seconds.
Low priority alarms are indicated by a message displayed on a steady yellow background and a sound signal made up of 2 beeps every 20 seconds.
The alarm screens offer two buttons:
▪ press the OK button for 2 seconds to acknowledge the alarm,
▪ press the MUTE button for 2 seconds to switch off the sound alarm for 2 minutes. The icon
is displayed in the bar at the top of the screen.