Cleaning and disinfection of the pump
Take the opportunity of a replacement of the SO
FILL, for example, to clean the SO
CONNECT PID pump. To do so, use a soft dry cloth to
clean the outside of the pump.
Disinfection of the SO
CONNECT PID pump is to achieve as follows:
Make sure that no SO
FILL syringe is connected to the pump. If necessary, remove it.
Clean coarse dirt with a damp cloth.
Turn on the screen.
Set up a new infusion with e 50 ml SO
FILL syringe and a volume of 1 ml.
Wait for the pusher to come into place.
Once the pusher is in position, remove the battery.
Disinfect the pusher, the pump housing, the battery compartment and its cover with a cloth moistened with Mikrozid AF or Puresept
AF.
Wait until the disinfectant solution has dried completely before handling the SO
CONNECT PID
pump.
Warning
Never clean your
SO
CONNECT PID
pump during an infusion. Cleaning could lead to a manipulating error and cause serious
injury.
Precautions
Do not use household or industrial cleaners, chemicals, solvents, bleach, scouring pads or sharp instruments to clean the
SO
CONNECT PID
pump.
Never put your
SO
CONNECT PID
pump in the dishwasher.
Always remove the
SO
FILL syringe
and remove the battery while cleaning.
Take care not to press too hard on the screen of the
SO
CONNECT PID
pump while cleaning it, so as to avoid damaging it.
The
SO
CONNECT PID
pump must not be sterilised.
Do not use cleaning and disinfecting products other than those recommended in this support. Non-recommended products may
damage the pump.